Sterilization and monthly spore testing in Iowa

How must an Iowa tattoo studio sterilize, and how often must it spore test?

Sterilize on site, and clean first

Iowa requires that every instrument needing sterilization be sterilized on site. Tubes, tips and grips that are not sterile, single-patron or disposable are cleaned first with detergent, following the maker's directions.

Then they go into a steam or dry-heat sterilizer before use on another person. Steam is preferred. The build-out guide covers where the sterilizer sits.

All instruments needing sterilization are sterilized on site, after being physically cleaned with a detergent according to manufacturers' recommendations. — Iowa Administrative Code 481, Chapter 492, via Iowa Legislature, retrieved 2026-10-01

The cycle settings the rule names

Steam sterilization runs at 250 degrees Fahrenheit (121 Celsius) for at least 15 minutes, at a minimum of 15 pounds per square inch.

Dry-heat sterilization runs at 350 degrees Fahrenheit (170 Celsius) for one hour. The rule says steam is preferred.

Read the sterilizer's manual against those numbers. A machine that cannot hold them does not meet the rule.

Iowa sterilization settings under 481 IAC 492.5
MethodTemperatureTime and pressure
Steam250 F (121 C)15 minutes at 15 psi minimum
Dry heat350 F (170 C)One hour

250 F, 15 minutesSteam sterilization is at 250 degrees Fahrenheit for 15 minutes at a minimum of 15 pounds per square inch, and dry-heat sterilization is at 350 degrees Fahrenheit for one hour. — Iowa Administrative Code 481, Chapter 492, via Iowa Legislature, retrieved 2026-10-01

Pouches and the 30-day date

After sterilizing, instruments are placed in closed pouches. Each pouch is dated effective for 30 days.

After 30 days the instruments are resterilized and the pouch is redated. Do not use an expired pouch for a client.

Tape a date label on every pouch the day it comes out of the sterilizer, and check the oldest first.

30 daysSterilized instruments are placed in closed pouches dated effective for 30 days, after which they are resterilized and the pouches redated. — Iowa Administrative Code 481, Chapter 492, via Iowa Legislature, retrieved 2026-10-01

The monthly spore test

Sterilizers are monitored monthly for spores of Bacillus subtilis. Results are kept for three years.

Run the test on a fixed day each month and file the result with the date. A shop that cannot show twelve months of results has no proof the sterilizer works.

Written procedures for a positive test must also be on file and in use. The records page lists the full retention set.

monthlySterilizers are monitored monthly for spores of Bacillus subtilis, and records of results are maintained for three years, along with written procedures for positive tests. — Iowa Administrative Code 481, Chapter 492, via Iowa Legislature, retrieved 2026-10-01

What a positive spore test triggers

A positive test pulls the sterilizer out of service at once. Everything processed from the cycle with the positive indicator until the next good result is treated as nonsterile and reprocessed before use.

The owner must tell the inspection agency within 24 hours of getting the result. The sterilizer may return to service only after one negative spore test.

Keep a second sterilizer or a supply of sterile single-use items, so a failure does not close the shop.

24 hoursA sterilizer with a positive spore test is removed from service at once, the owner notifies the inspection agency within 24 hours, and one negative test precedes reuse. — Iowa Administrative Code 481, Chapter 492, via Iowa Legislature, retrieved 2026-10-01

Single-use items and machine-side hygiene

Needles, scalpel blades, stencils, ink cups, gloves, gauze, razors and similar items are single use. They are used once and thrown away. Cloth towels and linens are not single use and are barred.

After each tattoo the artist cleans the machine head with an acceptable disinfectant and sprays surface disinfectant over the work area before the next setup. Items touched by gloved hands carry barrier films.

Used needles go straight into the red sharps container.

Single-use items include tattoo needles, scalpel blades, stencils, ink cups and protective gloves, and the artist cleans the machine head and sprays the work area after each tattoo. — Iowa Administrative Code 481, Chapter 492, via Iowa Legislature, retrieved 2026-10-01

Questions

Can I skip the autoclave by buying pre-sterilized supplies?

The rule requires instruments needing sterilization to be sterilized on site. Pre-sterilized single-use items need no sterilizing, but anything reusable does. Ask the inspector how your setup is classified before you buy.

What is the spore test organism?

The rule names spores of Bacillus subtilis and sets no test provider. Whichever method you choose, keep the dated result for three years.

Who do I call about a positive test?

The inspection agency, within 24 hours. That is the department or, in some places, a city, county or district board of health that inspects under an agreement with the department.